Magneto Mass-spectrometerBasic characteristics
Range of measurements of the volume fraction of components
The principle of operation of magnetic statistical mass-spectrometers is based on spatial division of ions of different masses, in their motion in a homogeneous magnetic field. In magnetic static mass-analyzers constant magnetic and electric fields are used. You can see the scheme of mass-analyzers in the figure. Such mass-analyzers possess properties of double focusing and provides resolution (100-500) even at the small sizes of the device. A source of electrons is a straight line cathode (1a.) Electrons, being accelerated up to energy (20-90) eV, ionize neutral molecules of gas test in an ionic source (1). When leaving the source ions are accelerated. Ions are divided by their masses in a homogeneous magnet field (3) which provides the resolution of mass-analyzer together with the electrostatic cylindrical energy analyzer (2). The rimer occurs due to the change of ionic energy in an ionic source. An electronic multiplier is used as a detector. |
|
||||||||||||||||||||||||||||||||||||||||||
|
|